Double Reed type frequencymeters

2HLC96 56-64Hz

2HLC96 56-64Hz, Frequencymeter

Code: M129320090000

  • Accuracy: 0,5
  • Measure: 56...64 Hz
  • Frequency (Hz): 56...64, 17 láminas
  • Modules: 96x96

INFORMATION

Description

  • Does not need an auxiliary power supply
  • DIN box with dimensions 96 and 144 mm
  • Class 1.5
  • Double scale

Applications

For the measurement and easy comparison of frequencies in alternating current circuits coming from generators or between the network and generator, when connected in parallel. The measurement is independent of the wave shape. For applications in severe environmental and physical conditions.

SPECIFICATIONS

Mechanical characteristics
Size (mm) width x height x depth
96 x 96 x 82.9 (mm)
Environmental characteristics
Protection class
Panel: IP 52 (Front), IP 00 (Terminals)
Storage temperature
-25…+40 °C
Working temperature
+10 … +30 °C
Standards
Certifications
UL
Electrical safety, Maximum height (m)
2000
Standards
BS 89, UNE-EN 60051, IEC 144, UL94, DIN 43780, IEC 51, UNE 21318
Voltage measurement circuit
Consumption
1 … 3,6 VA
Sampling frequency
47 … 53 Hz
Nominal voltage
230 V ~
Electrical characteristics
Insulation voltage, circuit
2kV a 50 Hz < 1min entre mecanismo y caja
Measurement accuracy
Accuracy
0,5 % FE

DIMENSIÓN

VÍDEOS

DESCARGAS

Datasheet

ACCESORIOS